Integrated Rentals part of Avante Health Solutions

Visit our new website at https://avantehs.com/rental

Refurbished - Respironics Novametrix Tidal Wave 710Sp / 715Sp Monitor

  • Web ID: 512
  • Condition: Retired

0 items are in your cart. View cart now.

This product is either very difficult to find or no longer available through Integrated, please see the list of related items below or ask your equipment expert for advice on a similar available product.

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (877) 907-0878

Description

FEATURES:

  • Built in Pulse Oximeter
  • Hand-held Mainstream and Sidestream Capnograph
  • 30-minute on-screen graphic trend available
  • Visual and audible ƒ¢¢€š¬…€œno respiration” alert
  • User replaceable disposable and rechargeable battery packs eliminate the need to wait while batteries are charging (standard AA lithium: 5.5 hours of life)
  • No calibration required for CO2


CAPNOGRAPH


Principle of Operation
Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.

Initialization Time
Capnogram within 15 seconds, full specifications within 60 seconds.

Calibration
No calibration required.

CAPNOSTAT® Mainstream CO2 Sensor
Size: 3.30 ƒÆ’— 4.24 ƒÆ’— 2.16 cm (1.30 ƒÆ’— 1.67 ƒÆ’— 0.85 in.).
Weight: 18 grams (0.64 oz.), cable excluded
Shock Resistance: Withstands repeated six foot drops to a hard surface.

CO2 (Carbon Dioxide)
Range: 0 - 150 mmHg
Accuracy:
‚‚ ‚±2 mmHg (for 0 - 40 mmHg),
‚‚ ‚±5% of reading (for 41 - 70 mmHg),
‚‚ ‚±8% of reading (for 71 - 150 mmHg)
Resolution: 1 mmHg
Response Time: 60 ms
Units: mmHg, kPa, or %
Compensations: N2O & O2 (selectable), Pb (automatic)
Capnogram: Selectable sweep speed

Respiratory Rate
Accuracy: ‚±1 br/min
Range: 0-150 br/min
Resolution: 1 br/min

Sampling Flow Rate (Model 715 only)
180 ml/min (for non-intubated patients)
Response time is less than 200 ms at 180 ml/min flow

PULSE OXIMETER

Principle of Operation
Red/Infrared absorption.

SpO2 (Oxygen Saturation)
Range: 0-100%
Accuracy: ‚±2% SpO2 (for 80-100% SpO2)
(1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
Resolution: 1%
Averaging: Menu selectable 2 or 8 seconds

Pulse Rate
Range: 30-250 bpm
Accuracy: ‚±1% of full scale
Resolution: 1 bpm
Averaging: 8 seconds

Sensors
Reusable Y-Sensor (can be sterilized and used with all patient populations);
reusable adult finger sensor

GENERAL

Alerts
Limits:Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout.
Audio: Adjustable volume, two minute silence or off (LED indicators)
Visual: On-screen and LED

Display
Type: Dot matrix, backlit (LCD)
Size: 2.2 ƒÆ’— 1.1 inches

Trending Capabilities
Memory: 24 hr., battery backed
Format: 30 minute on-screen or printed in trend or tabular mode.
Download data to personal computer for use with NovaCARD software

Communication Output
RS232 to Seiko DPU-414 printer, NovaCARD software or computer

Physical
Size (height, width, depth): 20 ƒÆ’— 9.5 ƒÆ’— 4.8 cm (8 ƒÆ’— 3.75 ƒÆ’— 1.875 in.).
Weight: 793 grams (28 oz.).

Electrical
External Supply (optional)
Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
Battery: Nickel-Metal-Hydride (NiMH): 4.5 hour life
Battery: Standard AA lithium: 5.5 hour life

Environmental
Operating temperature: 0 to +40 ‚°C (32-104 ‚°F)
Operating humidity: 0 to 90% RH non-condensing.

24-hour nationwide delivery is available*